legendsemi

LHE7909-4F

The LHE7909 series consists of 4-channel, 6-channel and 8-channel low-noise 24-bit synchronous sampling Δ-Σ analog-to-digital converters (ADCs). This series integrates programmable gain amplifiers (PGAs), an internal voltage reference and an on-chip oscillator. The LHE7909 integrates all conventional functions required for extracranial electroencephalogram (EEG) and electrocardiogram (ECG) applications. Benefiting from high integration and excellent performance, the LHE7909 supports the construction of scalable medical instrument systems with greatly reduced device size, power consumption and overall system cost. Each channel of the LHE7909 is equipped with a flexible input multiplexer, which can be independently connected to internally generated signals for self-testing, temperature measurement and lead-off detection. The LHE7909 operates at a data rate ranging from 250 SPS to 32 kSPS. On-chip lead-off detection can be realized by excitation current sources. Multiple LHE7909 devices can be cascaded in daisy-chain topology for high-channel-count systems. The LHE7909 is packaged in LQFP-64, with an operating temperature range of –40°C to +85°C.
LHE7909-4F

Product Features

• Up to 8 low-noise programmable gain amplifiers (PGAs) and 8 high-resolution synchronous sampling analog-to-digital converters (ADCs)

• Input-referred noise:

– 0.8 μVPP @ 250 SPS, Gain = 24

– 1 μVPP @ 250 SPS, Gain = 6

• Input bias current: 500 pA

• Data rate: 250 SPS to 32 kSPS

• Common-mode Rejection Ratio (CMRR): −120 dB

• Programmable gain options: 1, 2, 3, 4, 6, 8, 12, 24

• Unipolar or bipolar power supply:

– Analog supply (AVDD): 3 V or 4 V to 5.5 V

– Digital supply (DVDD): 1.8 V to 5.5 V

• Integrated bias drive amplifier, continuous lead-off detection and test signal generation

• On-chip oscillator

• Internal or external voltage reference available

• Flexible power-saving and standby modes

• Serial interface compatible with Serial Peripheral Interface (SPI)

• Operating temperature range: −40°C to +85°C

Application Scenario

Medical devices, including:

– Electroencephalogram (EEG) research

– Fetal electrocardiogram (Fetal ECG)

– Sleep research monitors

– Bispectral Index (BIS) monitoring

– Evoked Auditory Potential (EAP) testing